An H-module linear actuator for medical equipment applications

Xiao Liu, Keyuan Wu, yunyue ye, Kaiyuan Lu

Publikation: Bidrag til tidsskriftKonferenceartikel i tidsskriftForskningpeer review

22 Citationer (Scopus)

Abstract

An H-module linear actuator (HMLA) is proposed in this paper for medical equipment applications. Compared to the existing linear actuators used in medical equipment, the proposed H-module linear actuator has much lower normal force, which makes use of an additional air-suspension system unnecessary. The working principle of the HMLA is illustrated in detail in this paper, and its tangential and normal force characteristics are demonstrated by using finite element analysis (FEA). It is proved that the normal force of the HMLA can be greatly reduced without sacrificing the useful tangential force component.

OriginalsprogEngelsk
TidsskriftJournal of Applied Physics
Vol/bind111
Udgave nummer7
Antal sider3
ISSN0021-8979
DOI
StatusUdgivet - 2012
Begivenhed56th annual conference on magnetism and magnetic materials - , USA
Varighed: 30 okt. 2011 → …

Konference

Konference56th annual conference on magnetism and magnetic materials
Land/OmrådeUSA
Periode30/10/2011 → …

Fingeraftryk

Dyk ned i forskningsemnerne om 'An H-module linear actuator for medical equipment applications'. Sammen danner de et unikt fingeraftryk.