Analytical and between-subject variation of thrombin generation measured by calibrated automated thrombography on plasma samples

Anne F. Kristensen*, Søren R. Kristensen, Ursula Falkmer, Anna-Marie B. Münster, Shona Pedersen

*Kontaktforfatter

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

1 Citationer (Scopus)

Abstract

BACKGROUND: The Calibrated Automated Thrombography (CAT) is an in vitro thrombin generation (TG) assay that holds promise as a valuable tool within clinical diagnostics. However, the technique has a considerable analytical variation, and we therefore, investigated the analytical and between-subject variation of CAT systematically. Moreover, we assess the application of an internal standard for normalization to diminish variation.

METHODS: 20 healthy volunteers donated one blood sample which was subsequently centrifuged, aliquoted and stored at -80 °C prior to analysis. The analytical variation was determined on eight runs, where plasma from the same seven volunteers was processed in triplicates, and for the between-subject variation, TG analysis was performed on plasma from all 20 volunteers. The trigger reagents used for the TG assays included both PPP reagent containing 5 pM tissue factor (TF) and PPPlow with 1 pM TF. Plasma, drawn from a single donor, was applied to all plates as an internal standard for each TG analysis, which subsequently was used for normalization.

RESULTS: The total analytical variation for TG analysis performed with PPPlow reagent is 3-14% and 9-13% for PPP reagent. This variation can be minimally reduced by using an internal standard but mainly for ETP (endogenous thrombin potential). The between-subject variation is higher when using PPPlow than PPP and this variation is considerable higher than the analytical variation.

CONCLUSION: TG has a rather high inherent analytical variation but considerable lower than the between-subject variation when using PPPlow as reagent.

OriginalsprogEngelsk
TidsskriftScandinavian Journal of Clinical & Laboratory Investigation
Vol/bind78
Udgave nummer3
Sider (fra-til)175-179
Antal sider5
ISSN0036-5513
DOI
StatusUdgivet - 2018

Fingeraftryk

Dyk ned i forskningsemnerne om 'Analytical and between-subject variation of thrombin generation measured by calibrated automated thrombography on plasma samples'. Sammen danner de et unikt fingeraftryk.

Citationsformater