Association between implantable cardioverter-defibrillator therapy and different lead positions in patients with cardiac resynchronization therapy

Mads Brix Kronborg, Jens Brock Johansen, Jens Haarbo, Sam Riahi, Berit Thornvig Philbert, Ole Dan Jørgensen, Jens Cosedis Nielsen

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

4 Citationer (Scopus)

Abstract

Aims: To evaluate the impact of different right and left ventricular lead positions (RV-LP and LV-LP) on the risk of therapy for ventricular tachycardia/ventricular fibrillation in patients with a cardiac resynchronization therapy device (CRT-D).

Methods and results: We performed a large nationwide cohort study on patients in Denmark receiving a CRT-D device from 2008 to 2012 from the Danish Pacemaker and implantable cardioverter defibrillator (ICD) registry. Lead positions were registered during the implantation and categorized as anterior/lateral/posterior and basal/mid-ventricular/apical for the LV-LP, and as apical/non-apical for the RV-LP. Appropriate and inappropriate therapies were registered during follow-up via remote monitoring or at device interrogations. Time to event was summarized with Kaplan-Meier plots, and competed risk regression analysis was used to calculate adjusted hazard ratios (aHR) with 95% confidence intervals (CI). Following variables were included in the analysis: gender, age, heart failure aetiology, New York heart association class, left ventricular ejection fraction, QRS duration, indication (secondary or primary prophylactic), RV-LP, LV-LP, and antiarrhythmic therapy. We included 1643 patients [mean age 68 (±10) years, 1343 (83%) men]. After a mean of 2.0 years, 322 (20%) patients received appropriate and 66 (4%) patients received inappropriate therapy. The aHR for appropriate therapy with a non-apical RV-LP was 0.70 95% CI (0.55-0.87, P = 0.002) as compared with an apical. We observed no significant association between appropriate therapy and LV-LP in left anterior oblique or right anterior oblique views or inappropriate therapy between any lead positions.

Conclusion: An apical RV-LP is associated with an increased risk of appropriate therapy for ventricular tachyarrhythmia in patients with a CRT-D device.

OriginalsprogEngelsk
TidsskriftEuropace
Vol/bind20
Udgave nummer9
Sider (fra-til)E133-E139
Antal sider7
ISSN1099-5129
DOI
StatusUdgivet - 2018

Fingeraftryk

Dyk ned i forskningsemnerne om 'Association between implantable cardioverter-defibrillator therapy and different lead positions in patients with cardiac resynchronization therapy'. Sammen danner de et unikt fingeraftryk.

Citationsformater