Analytical and between-subject variation of thrombin generation measured by calibrated automated thrombography on plasma samples

Anne F. Kristensen*, Søren R. Kristensen, Ursula Falkmer, Anna-Marie B. Münster, Shona Pedersen

*Corresponding author for this work

Research output: Contribution to journalJournal articleResearchpeer-review

1 Citation (Scopus)

Abstract

BACKGROUND: The Calibrated Automated Thrombography (CAT) is an in vitro thrombin generation (TG) assay that holds promise as a valuable tool within clinical diagnostics. However, the technique has a considerable analytical variation, and we therefore, investigated the analytical and between-subject variation of CAT systematically. Moreover, we assess the application of an internal standard for normalization to diminish variation.

METHODS: 20 healthy volunteers donated one blood sample which was subsequently centrifuged, aliquoted and stored at -80 °C prior to analysis. The analytical variation was determined on eight runs, where plasma from the same seven volunteers was processed in triplicates, and for the between-subject variation, TG analysis was performed on plasma from all 20 volunteers. The trigger reagents used for the TG assays included both PPP reagent containing 5 pM tissue factor (TF) and PPPlow with 1 pM TF. Plasma, drawn from a single donor, was applied to all plates as an internal standard for each TG analysis, which subsequently was used for normalization.

RESULTS: The total analytical variation for TG analysis performed with PPPlow reagent is 3-14% and 9-13% for PPP reagent. This variation can be minimally reduced by using an internal standard but mainly for ETP (endogenous thrombin potential). The between-subject variation is higher when using PPPlow than PPP and this variation is considerable higher than the analytical variation.

CONCLUSION: TG has a rather high inherent analytical variation but considerable lower than the between-subject variation when using PPPlow as reagent.

Original languageEnglish
JournalScandinavian Journal of Clinical & Laboratory Investigation
Volume78
Issue number3
Pages (from-to)175-179
Number of pages5
ISSN0036-5513
DOIs
Publication statusPublished - 2018

Keywords

  • Journal Article
  • between-subject variation
  • thrombin
  • blood coagulation
  • blood coagulation test
  • thrombin generation
  • Analytical variation
  • Reproducibility of Results
  • Humans
  • Middle Aged
  • Thrombin/analysis
  • Male
  • Calibration
  • Healthy Volunteers
  • Thrombelastography/methods
  • Analysis of Variance
  • Blood Coagulation Tests
  • Adult
  • Female
  • Thromboplastin/chemistry
  • Observer Variation

Fingerprint

Dive into the research topics of 'Analytical and between-subject variation of thrombin generation measured by calibrated automated thrombography on plasma samples'. Together they form a unique fingerprint.

Cite this