Abstract

BACKGROUND: The present paper describes the design and evaluation of an automated version of the Modified Jebsen Test of Hand Function (MJT) based on the Microsoft Kinect sensor.

METHODS: The MJT was administered twice to 11 chronic stroke subjects with varying degrees of hand function deficits. The test times of the MJT were evaluated manually by a therapist using a stopwatch, and automatically using the Microsoft Kinect sensor. The ground truth times were assessed based on inspection of the video-recordings. The agreement between the methods was evaluated along with the test-retest performance.

RESULTS: The results from Bland-Altman analysis showed better agreement between the ground truth times and the automatic MJT time evaluations compared to the agreement between the ground truth times and the times estimated by the therapist. The results from the test-retest performance showed that the subjects significantly improved their performance in several subtests of the MJT, indicating a practice effect.

CONCLUSIONS: The results from the test showed that the Kinect can be used for automating the MJT.

Original languageEnglish
Article number38
JournalJournal of NeuroEngineering and Rehabilitation
Volume14
Number of pages12
ISSN1743-0003
DOIs
Publication statusPublished - 2017