Radiotherapy in patients with pacemakers and implantable cardioverter defibrillators: a literature review

Research output: Contribution to journalReview articlepeer-review

63 Citations (Scopus)

Abstract

An increasing number of patients with implantable cardiac rhythm devices undergo radiotherapy (RT) for cancer and are thereby exposed to the risk of device failure. Current safety recommendations seem to have limitations by not accounting for the risk of pacemakers and implantable cardioverter defibrillators malfunctioning at low radiation doses. Besides scant knowledge about optimal safety measures, only little is known about the exact prevalence of patients with devices undergoing RT. In this review, we provide a short overview of the principles of RT and present the current evidence on the predictors and mechanisms of device malfunctions during RT. We also summarize practical recommendations from recent publications and from the industry. Strongly associated with beam energy of photon RT, device malfunctions occur at ∼3% of RT courses, posing a substantial issue in clinical practice. Malfunctions described in the literature typically consist of transient software disturbances and only seldom manifest as a permanent damage of the device. Through close cooperation between cardiologists and oncologists, a tailored individualized approach might be necessary in this patient group in waiting time for updated international guidelines in the field.

Original languageEnglish
JournalEuropace
Volume18
Issue number4
Pages (from-to)479-491
Number of pages13
ISSN1099-5129
DOIs
Publication statusPublished - 2016

Fingerprint

Dive into the research topics of 'Radiotherapy in patients with pacemakers and implantable cardioverter defibrillators: a literature review'. Together they form a unique fingerprint.

Cite this