Abstract
Introduction: Internal rigid fixation techniques are commonly used to treat maxillofacial fractures by stabilizing bone segments using titanium plates and screws. The current study aimed to compare the biomechanical stability of two maxillofacial screws with different tip designs in vivo. Materials and Methods: Six male rabbits were randomly divided into two experimental periods of 0 and 4 weeks. Under general anesthesia, the screws were randomly placed in the tibia bone on both sides of each animal. The pullout test was conducted using a Sentam test device. Data were statistically analyzed using ANOVA. Results: The average insertion torque for non-self-tapping and self-tapping screws amounted to 4.2±1.7 and 4.8±1.4 Newton/centimeter, respectively (P-value >0.05). The calculated measures for the pullout test demonstrated a significant increase of secondary stability after 4 weeks in comparison to 0 week (P-value0.05). Conclusion: No significant differences were demonstrated among self-tapping or non-self-tapping maxillofacial screws.