Back Scatter Interferometry (BSI) has been proposed to be a highly sensitive and versatile refractive index sensor usable for analytical detection of biomarker and protein interactions in solution. However the existing literature on BSI lacks a physical explanation of why protein interactions in general should contribute to the BSI signal. We have established a BSI system to investigate this subject in further detail. We contribute with a thorough analysis of the robustness of the sensor including unwanted contributions to the interferometric signal caused by temperature variation and dissolved gasses. We report a limit of the effective minimum detectability of refractive index at the 10(-7) level. Long term stability was examined by simultaneously monitoring the temperature inside the capillary revealing an average drift of 2.0 × 10(-7) per hour. Finally we show that measurements on protein A incubated with immunoglobulin G do not result in a signal that can be attributed to binding affinities as otherwise claimed in literature.
Originalsprog | Engelsk |
---|---|
Tidsskrift | The Analyst |
Vol/bind | 140 |
Udgave nummer | 3 |
Sider (fra-til) | 895-901 |
Antal sider | 7 |
ISSN | 0003-2654 |
DOI | |
Status | Udgivet - 2015 |