T1 - Evaluation of decision support to wean patients from mechanical ventilation in intensive care
T2 - a prospective study reporting clinical and physiological outcomes
AU - Vizcaychipi, Marcela P
AU - Karbing, Dan S
AU - Martins, Laura
AU - Gupta, Amandeep
AU - Moreno-Cuesta, Jeronimo
AU - Naik, Manu
AU - Welters, Ingeborg
AU - Singh, Suveer
AU - Randell, Georgina
AU - Osman, Leyla
AU - Rees, Stephen E
N1 - © 2024. The Author(s).
PY - 2025/4
Y1 - 2025/4
N2 - This study investigated the clinical and physiological response to use of the BEACON Caresystem, a bedside open-loop decision support system providing advice to guide clinicians when weaning patients from invasive mechanical ventilation. Multicenter prospective study conducted in five adult intensive care units in the UK. Following screening and assent, intubated patients mechanically ventilated for > 24 h were randomized to intervention or usual care. Intervention consisted of application of the BEACON Caresystem’s advice on tidal volume/inspiratory pressure, inspired oxygen, respiratory rate and PEEP. Usual care was defined as local clinical practice. The primary outcome was duration of mechanical ventilation. Secondary outcomes quantified prolonged intubation and survival; adverse events; ventilator settings and physiological state; time spent in ventilator modes; links to other therapy; the frequency of advice utilization and time spent outside normal physiological limits. The study was terminated early with a total of 112 patients included. Fifty-four were randomised to the intervention arm and fifty-eight to usual care. The study was underpowered and no significant differences were seen in duration of mechanical ventilation (p = 0.773), prolonged intubation or survival. Intervention arm patients had lower rates of adverse events (p = 0.016), including fewer hypoxaemic events (p = 0.008) and lower values of PEEP (p = 0.030) and tidal volume (p = 0.042). Values of peak inspiratory pressure and pressure support were reduced but at the boarder of statistical significance (p = 0.104, p = 0.093, respectively). No differences were seen for time in ventilator mode or other therapy. Advice presented by the decision support system was applied at the beside an average of 88% of occasions, with a significantly increased number of changes only in inspired oxygen fraction. No significant differences were seen in time spent outside physiological limits. This study investigated the use of the BEACON Caresystem, an open loop clinical decision support system providing advice on ventilator settings. It was terminated early, with no significant difference shown in duration of mechanical ventilation, the primary outcome. Application of advice indicated potential for fewer adverse events and improved physiological status. (Trial registration ClinicalTrials.gov under NCT03249623.
AB - This study investigated the clinical and physiological response to use of the BEACON Caresystem, a bedside open-loop decision support system providing advice to guide clinicians when weaning patients from invasive mechanical ventilation. Multicenter prospective study conducted in five adult intensive care units in the UK. Following screening and assent, intubated patients mechanically ventilated for > 24 h were randomized to intervention or usual care. Intervention consisted of application of the BEACON Caresystem’s advice on tidal volume/inspiratory pressure, inspired oxygen, respiratory rate and PEEP. Usual care was defined as local clinical practice. The primary outcome was duration of mechanical ventilation. Secondary outcomes quantified prolonged intubation and survival; adverse events; ventilator settings and physiological state; time spent in ventilator modes; links to other therapy; the frequency of advice utilization and time spent outside normal physiological limits. The study was terminated early with a total of 112 patients included. Fifty-four were randomised to the intervention arm and fifty-eight to usual care. The study was underpowered and no significant differences were seen in duration of mechanical ventilation (p = 0.773), prolonged intubation or survival. Intervention arm patients had lower rates of adverse events (p = 0.016), including fewer hypoxaemic events (p = 0.008) and lower values of PEEP (p = 0.030) and tidal volume (p = 0.042). Values of peak inspiratory pressure and pressure support were reduced but at the boarder of statistical significance (p = 0.104, p = 0.093, respectively). No differences were seen for time in ventilator mode or other therapy. Advice presented by the decision support system was applied at the beside an average of 88% of occasions, with a significantly increased number of changes only in inspired oxygen fraction. No significant differences were seen in time spent outside physiological limits. This study investigated the use of the BEACON Caresystem, an open loop clinical decision support system providing advice on ventilator settings. It was terminated early, with no significant difference shown in duration of mechanical ventilation, the primary outcome. Application of advice indicated potential for fewer adverse events and improved physiological status. (Trial registration ClinicalTrials.gov under NCT03249623.
KW - Decision support
KW - Mechanical ventilation
KW - Randomised control trial
UR - http://www.scopus.com/inward/record.url?scp=85208784123&partnerID=8YFLogxK
U2 - 10.1007/s10877-024-01231-5
DO - 10.1007/s10877-024-01231-5
M3 - Journal article
C2 - 39520605
SN - 1387-1307
VL - 39
SP - 393
EP - 404
JO - Journal of Clinical Monitoring and Computing
JF - Journal of Clinical Monitoring and Computing
IS - 2
M1 - e042145
ER -