Simple Post-Processing of Continuous Glucose Monitoring Measurements Improves Endpoints in Clinical Trials

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

4 Citationer (Scopus)

Abstract

BACKGROUND: Continuous glucose monitoring (CGM) is a powerful tool to be considered both in clinical practice and clinical trials. However, CGM has been criticized for being inaccurate for many reasons including a physiological delay. This study sought to investigate the current delay issue and propose a simple post-processing procedure.

METHOD: More than a million hours of the Dexcom G4 CGM from 472 subjects investigated in a state-of-the-art clinical trial were analyzed by time shifting the CGM measurements and comparing them to plasma glucose (PG) measurements. The resultant CGM measurements were then assessed in relation to real-world clinical research endpoints.

RESULTS: A CGM time shift of -9 minutes was optimal and reduced mean absolute relative difference (MARD) statistically significantly with 1.0% point. The MARD reduction resulted in better clinical research endpoints of hypoglycemia and postprandial glucose increments.

CONCLUSIONS: The delay in CGM is still an issue. The delay in this study was identified to be 9 minutes compared to PG. With a simple post-processing approach of time shifting the CGM measurements with -9 minutes, it was possible to obtain a statistically significantly lower MARD and subsequently obtain clinical research endpoints of improved validity.