Heart rate variability density analysis (Dyx) for identification of appropriate implantable cardioverter defibrillator recipients among elderly patients with acute myocardial infarction and left ventricular systolic dysfunction

Rikke Mørch Jørgensen, Jacob Levitan, Zohar Halevi, Natalia Puzanov, Steen Zabell Abildstrøm, Marc D Messier, Heikki V Huikuri, Jens Haarbo, Poul Erik Bloch Thomsen, Christian Jons, on behalf of the CARISMA investigators

Research output: Contribution to journalJournal articleResearchpeer-review

4 Citations (Scopus)

Abstract

AIMS: Dyx is a new heart rate variability (HRV) density analysis specifically designed to identify patients at high risk for malignant ventricular arrhythmias. The aim of this study was to test if Dyx can improve risk stratification for malignant ventricular tachyarrhythmias and to test if the previously identified cut-off can be reproduced.

METHODS AND RESULTS: This study included 248 patients from the CARISMA study with ejection fraction ≤40% after an acute myocardial infarction and an analysable 24 h Holter recording. All patients received an implantable cardiac monitor, which was used to diagnose the primary endpoint of near-fatal or fatal ventricular tachyarrhythmias likely preventable by an implantable cardioverter defibrillator (ICD), during a period of 2 years. A Dyx ≤ 1.96 was considered abnormal. The secondary endpoint was cardiovascular death. At enrolment 59 patients (24%) had a Dyx ≤ 1.96 and 20 experienced a primary endpoint. A Dyx ≤ 1.96 was associated with a significantly increased risk for malignant arrhythmias [hazards ratio (HR) = 4.36 (1.81-10.52), P = 0.001] and cardiovascular death [HR = 3.47 (1.38-8.74), P = 0.008]. Compared with important clinical risk parameters (age >70 years and QRS > 120 ms), Dyx ≤ 1.96 significantly added predictive value (P = 0.0066).

CONCLUSIONS: Dyx was a better predictor of ventricular tachyarrhythmias than the traditional measures of HRV and heart rate turbulence, particularly in the elderly. Dyx might be a useful tool for better selection of ICD candidates in the elderly population, since a normal Dyx in this group was associated with a very low risk for malignant ventricular arrhythmias. ClinicalTrials.gov Identifier NCT00145119.

Original languageEnglish
Article numbereuu394
JournalEuropace
Volume17
Issue number12
Pages (from-to)1848-1854
Number of pages7
ISSN1099-5129
DOIs
Publication statusPublished - 2015

Fingerprint

Dive into the research topics of 'Heart rate variability density analysis (Dyx) for identification of appropriate implantable cardioverter defibrillator recipients among elderly patients with acute myocardial infarction and left ventricular systolic dysfunction'. Together they form a unique fingerprint.

Cite this