Is a single late SPECT/CT based kidney 177Lu-dosimetry superior to hybrid dosimetry with sequential multiple time-point whole-body planar scans in combination with an early SPECT/CT?

Seval Beykan*, Johannes Tran-Gia, Svend Borup Jensen, Michael Lassmann

*Corresponding author for this work

Research output: Contribution to journalJournal articleResearchpeer-review

1 Citation (Scopus)
19 Downloads (Pure)

Abstract

PURPOSE: The aim is to assess the impact of different imaging-protocols on image-based kidney dosimetry in 177Lu labelled peptide receptor radiotherapies.

METHODS: Kidney data of five [177Lu]Lu-OPS201 injected pigs and a 3D printed phantom were used for comparing the absorbed doses and time-integrated activity coefficients calculated based on the following imaging-protocols: A-) multiple time-point SPECT/CTs, B-) multiple time-point planar scans in combination with one SPECT/CT, C-) single time-point SPECT/CT. In addition, the influence of late scan time-points on kidney dosimetry was investigated by sequentially eliminating scan data at > 100 h from the pig/phantom datasets for imaging-protocols A and B.

RESULTS: Compared to imaging-protocol A, absorbed doses based on imaging-protocols B and C (scans at > 24 h post-injection) were always lower (differences > 34%). The best agreement in absorbed dose was achieved by imaging-protocol C at ∼ 100 h post-injection (difference: 4%). Regarding the phantom/pig experiments, eliminating scan data at > 100 h post-injection increased the time-integrated activity coefficients calculated based on imaging-protocols A and B by up to 83%.

CONCLUSION: While imaging-protocol A is accurate if scans at >∼100 h are included, it is time-consuming. In addition to being time-consuming, imaging-protocol B shows high differences associated with organ-count overlay, a lack of accuracy concerning the geometric mean based 2D attenuation correction, and 2D background subtraction due to the inhomogeneous and time-varying background contributions. Our findings indicate that dosimetry based on imaging-protocol C, if appropriately performed, provides similar kidney absorbed doses compared to imaging-protocol A, while only a single scan time-point is necessary.

Original languageEnglish
JournalPhysica Medica
Volume100
Pages (from-to)39-50
Number of pages12
ISSN1120-1797
DOIs
Publication statusPublished - Aug 2022

Bibliographical note

Copyright © 2022 Associazione Italiana di Fisica Medica e Sanitaria. Published by Elsevier Ltd. All rights reserved.

Keywords

  • Hybrid planar/SPECT/CT dosimetry
  • Lu dosimetry
  • Multiple time-point SPECT/CT dosimetry
  • Single time-point SPECT/CT dosimetry
  • Animals
  • Kidney/diagnostic imaging
  • Swine
  • Tomography, Emission-Computed, Single-Photon
  • Single Photon Emission Computed Tomography Computed Tomography
  • Phantoms, Imaging
  • Radiometry/methods