Skip to main navigation Skip to search Skip to main content

Medical tongue piercing: development and evaluation of a surgical protocol and the perception of procedural discomfort of the participants

Research output: Contribution to journal › Journal article › Research › peer-review

Abstract

Background
A system providing disabled persons with control of various assistive devices with the tongue has been developed at Aalborg University in Denmark. The system requires an activation unit attached to the tongue with a small piercing. The aim of this study was to establish and evaluate a safe and tolerable procedure for medical tongue piercing and to evaluate the expected and perceived procedural discomfort.

Methods
Four tetraplegic subjects volunteered for the study. A surgical protocol for a safe insertion of a tongue barbell piercing was presented using sterilized instruments and piercing parts. Moreover, post-procedural observations of participant complications such as bleeding, edema, and infection were recorded. Finally, procedural discomforts were monitored by VAS scores of pain, changes in taste and speech as well as problems related to hitting the teeth.

Results
The piercings were all successfully inserted in less than 5 min and the pain level was moderate compared with oral injections. No bleeding, infection, embedding of the piercing, or tooth/gingival injuries were encountered; a moderate edema was found in one case without affecting the speech. In two cases the piercing rod later had to be replaced by a shorter rod, because participants complained that the rod hit their teeth. The replacements prevented further problems. Moreover, loosening of balls was encountered, which could be prevented with the addition of dental glue. No cases of swallowing or aspiration of the piercing parts were recorded.

Conclusions
The procedure proved simple, fast, and safe for insertion of tongue piercings for tetraplegic subjects in a clinical setting. The procedure represented several precautions in order to avoid risks in these susceptible participants with possible co-morbidity. No serious complications were encountered, and the procedure was found tolerable to the participants. The procedure may be used in future studies with tongue piercings being a prerequisite for similar systems, and this may include insertion in an out-patient setting.
Original languageEnglish
Article number44
JournalJournal of NeuroEngineering and Rehabilitation
Volume11
Issue number1
Number of pages11
ISSN1743-0003
DOIs
Publication statusPublished - 2014

Keywords

  • Tongue control
  • Assistive device
  • Piercing
  • Surgical tongue piercing technique
  • Procedural pain
  • Expected pain
  • Perceived pain
  • VAS

Fingerprint

Dive into the research topics of 'Medical tongue piercing: development and evaluation of a surgical protocol and the perception of procedural discomfort of the participants'. Together they form a unique fingerprint.

Cite this