NT-proBNP on Cobas h 232 in point-of-care testing: Performance in the primary health care versus in the hospital laboratory

Charlotte Gils*, Ramshanker Ramanathan, Torben Breindahl, Mette Brokner, Anne L. Christiansen, Øyvin Eng, Ida J. Hammer, Catherine B. Herrera, Aina Jansen, Eva C. Langsjøen, Elisabeth S. Løkkebo, Toril Osestad, Anne D. Schrøder, Lisa Walther

*Corresponding author for this work

Research output: Contribution to journalJournal articleResearchpeer-review

13 Citations (Scopus)

Abstract

Background. NT-proBNP may be useful for ruling out heart failure in primary health care. In this study we examined the analytical quality of NT-proBNP in primary health care on the Cobas h 232 point-of-care instrument compared with measurements performed in a hospital laboratory. Materials and methods. Blood samples requested for NT-proBNP were collected in primary health care (n = 95) and in a hospital laboratory (n = 107). NT-proBNP was measured on-site on Cobas h 232 instruments both in primary health care centres and at the hospital laboratory and all samples were also analyzed with a comparison method at the hospital. Precision, trueness, accuracy, and lot-variation were determined at different concentration levels and evaluated according to acceptance criteria. Furthermore user-friendliness was assessed by questionnaires. Results. For Cobas h 232 repeatability CV was 8.5-10.7% in the hospital setting and 5.3-10.0% in the primary health care and within the analytical quality specifications, but higher than with the comparison method (< 4%). NT-proBNP results obtained in primary health care were significantly higher than by the hospital comparison method (bias ranged from 14.3-23.7%), whereas there was no significant bias when Cobas h 232 was used in the hospital setting (bias ranged from - 4.9 to 7.0%). User-friendliness of Cobas h 232 was overall acceptable. Conclusion. Cobas h 232 point-of-care instrument for measurement of NT-proBNP performed satisfactorily with regard to precision, user-friendliness, and lot-variation. A decrease in NT-proBNP levels observed in samples transported to a central laboratory needs further attention and investigation.

Original languageEnglish
Book seriesScandinavian Journal of Clinical and Laboratory Investigation
Volume75
Issue number7
Pages (from-to)602-609
Number of pages8
ISSN0036-5513
DOIs
Publication statusPublished - 3 Oct 2015

Bibliographical note

Publisher Copyright:
© 2015 Informa Healthcare.

Keywords

  • biological markers
  • heart disease
  • Natriuretic peptide
  • point-of-care systems
  • validation studies

Fingerprint

Dive into the research topics of 'NT-proBNP on Cobas h 232 in point-of-care testing: Performance in the primary health care versus in the hospital laboratory'. Together they form a unique fingerprint.

Cite this