Should blood flow during cardiopulmonary bypass be individualized more than to body surface area?

Research output: Contribution to journalJournal articleResearchpeer-review

4 Citations (Scopus)

Abstract

Blood flow during cardiopulmonary bypass (CPB) is calculated on body surface area (BSA). Increasing comorbidity, age and weight of today's cardiac patients question this calculation as it may not reflect individual metabolic requirement. The hypothesis was that a measured cardiac index (CI) prior to normothermic CPB is a better estimate. A cross-over study, with random allocation to CPB blood flow for 20 minutes based on either a calculation (2.4 L/min/m(2)) or on CI, with a switch to the opposite flow for another 20 minutes, was performed. Twenty-two elective cardiac surgery patients with normal ventricular function were included. Effect parameters were cerebral oxygenation, mixed venous saturation and arterial lactate. CI varied from 1.9 to 3.1 L/min/m(2) (median 2.4 L/min/m(2)). No differences in effect parameters were seen. In conclusion, a CPB blood flow based on an individual estimate did not improve cerebral and systemic oxygenation compared to a blood flow based on BSA.
Original languageEnglish
JournalPerfusion
Volume26
Pages (from-to)45-50
Number of pages6
ISSN0267-6591
DOIs
Publication statusPublished - 2011
Externally publishedYes

Fingerprint

Dive into the research topics of 'Should blood flow during cardiopulmonary bypass be individualized more than to body surface area?'. Together they form a unique fingerprint.

Cite this