Adverse Design of Defibrillators: Turning Off the Machine When Trying to Shock.

Christian S Høyer, Erika F Christensen, Berit Eika

Research output: Contribution to journalJournal articleResearchpeer-review

10 Citations (Scopus)

Abstract

A recent publication demonstrated the possibility of erroneous operation of 2 widely used monitor-defibrillators and observed that the design of user interfaces might contribute to error during operation. During an ambulance simulation training exercise for 72 junior internal medicine physicians that called for defibrillation in the management of cardiac arrest, we observed that in 5 of 192 defibrillation attempts by the physicians, the monitor-defibrillator was inadvertently powered off. When the device is inadvertently powered off, recognition and subsequent steps to defibrillate delayed defibrillation an average of 24 seconds (range 14 to 32 seconds). Our analysis of the controls of this monitor-defibrillator found that the device could be powered off even if fully charged and ready to shock. Redesign of the equipment might prevent this inadvertent event.
Original languageEnglish
JournalAnnals of Emergency Medicine
Volume52
Issue number5
Pages (from-to)512-514
Number of pages3
ISSN0196-0644
DOIs
Publication statusPublished - 2008
Externally publishedYes

Fingerprint

Dive into the research topics of 'Adverse Design of Defibrillators: Turning Off the Machine When Trying to Shock.'. Together they form a unique fingerprint.

Cite this