Comparison of test results from two separate video head impulse test systems in a cohort of patients diagnosed with a unilateral vestibular schwannoma

Mathilde Aalling*, Regitze Kuhr Skals, Emil Riis Abrahamsen, Dan Dupont Hougaard

*Kontaktforfatter

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

3 Citationer (Scopus)

Abstract

PURPOSE: Video head impulse testing (vHIT) is a relatively new technology enabling evaluation of vestibular function. The aim of this study was to compare the test results from two separate vHIT systems in a group of patients diagnosed with a unilateral vestibular schwannoma (VS) with regards to sensitivity, specificity and inter-examiner differences.

METHODS: Forty-two patients were examined with two separate vHIT systems: EyeSeeCam® (system A) and ICS Impulse® (system B), by one of two examiners. All six semicircular canals (SCCs) were tested under standardized conditions, and strict criteria were set up for post-test interpretation.

RESULTS: With the majority of test parameters, the two test systems were in agreement. Vestibular deficits were found in 40.5% (system A) to 45% (system B) of patients with a VS on the tested side; corresponding to a positive predictive value (PPV) of 86.4% (system B) to 94.4% (system A). The specificity was 97.6% for system A and 92.9% for system B. An overall agreement between the two vHIT systems measured as kappa was computed to be 0.61. There were no significant inter-examiner differences. When testing the vertical SCCs, a tendency of too high mean gain values was seen with system A but not with system B.

CONCLUSION: In patients with unilateral VS, vHIT is a test with moderate sensitivity and high specificity in regard to identification of a vestibular deficit. There were no significant differences in test results between the two vHIT systems.

OriginalsprogEngelsk
TidsskriftEuropean Archives of Oto-Rhino-Laryngology
Vol/bind277
Udgave nummer11
Sider (fra-til)3185-3193
Antal sider9
ISSN0937-4477
DOI
StatusUdgivet - nov. 2020