Continuous cardiac output measured with a Swan-Ganz catheter reacts too slowly in animal experiments with sudden circulatory failure

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

2 Citationer (Scopus)
38 Downloads (Pure)

Abstract

BACKGROUND: In many animal experiments, it is vital to detect sudden changes in cardiac output (CO). This porcine study compared CO that was measured with a Swan-Ganz pulmonary catheter with the gold standard (which was a transit-time flow probe around the pulmonary artery) during interventions that caused hemodynamic instability.

METHODS: In one series, 7 pigs were exposed to sudden changes in CO. In another series, 9 pigs experienced more prolonged changes in CO. All the pigs had a Swan-Ganz catheter placed into the pulmonary artery and a flow probe around the pulmonary artery. Adrenaline infusion and controlled hemorrhage were used to increase and decrease CO, respectively. The measurements of CO before and after each intervention were compared for correlation, agreement, and the time delay that it took each method to detect at least a 30% change in CO. A Bland-Altman test was used to identify correlations and agreements between the methods.

RESULTS: In the first series, there was a delay of 5-7 min for the Swan Ganz catheter to register a 30% change in cardiac output, compared with the flow probe. However, during prolonged changes in CO in the second series, there was a good correlation between the 2 methods. Mixed venous oxygen saturation reacted faster to changes than did CO; both were measured via the Swan-Ganz catheter.

CONCLUSIONS: In many animal studies, the use of Swan-Ganz catheters is suitable; however, in experiments with sudden hemodynamic instability, the flow probe is the most advantageous method for measuring CO.

OriginalsprogEngelsk
TidsskriftAnimal models and experimental medicine
Vol/bind5
Udgave nummer3
Sider (fra-til)258-265
Antal sider8
ISSN2096-5451
DOI
StatusUdgivet - jun. 2022

Bibliografisk note

© 2022 The Authors. Animal Models and Experimental Medicine published by John Wiley & Sons Australia, Ltd on behalf of The Chinese Association for Laboratory Animal Sciences.

Fingeraftryk

Dyk ned i forskningsemnerne om 'Continuous cardiac output measured with a Swan-Ganz catheter reacts too slowly in animal experiments with sudden circulatory failure'. Sammen danner de et unikt fingeraftryk.

Citationsformater