NT-proBNP on Cobas h 232 in point-of-care testing: Performance in the primary health care versus in the hospital laboratory

Charlotte Gils*, Ramshanker Ramanathan, Torben Breindahl, Mette Brokner, Anne L. Christiansen, Øyvin Eng, Ida J. Hammer, Catherine B. Herrera, Aina Jansen, Eva C. Langsjøen, Elisabeth S. Løkkebo, Toril Osestad, Anne D. Schrøder, Lisa Walther

*Kontaktforfatter

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

12 Citationer (Scopus)

Abstract

Background. NT-proBNP may be useful for ruling out heart failure in primary health care. In this study we examined the analytical quality of NT-proBNP in primary health care on the Cobas h 232 point-of-care instrument compared with measurements performed in a hospital laboratory. Materials and methods. Blood samples requested for NT-proBNP were collected in primary health care (n = 95) and in a hospital laboratory (n = 107). NT-proBNP was measured on-site on Cobas h 232 instruments both in primary health care centres and at the hospital laboratory and all samples were also analyzed with a comparison method at the hospital. Precision, trueness, accuracy, and lot-variation were determined at different concentration levels and evaluated according to acceptance criteria. Furthermore user-friendliness was assessed by questionnaires. Results. For Cobas h 232 repeatability CV was 8.5-10.7% in the hospital setting and 5.3-10.0% in the primary health care and within the analytical quality specifications, but higher than with the comparison method (< 4%). NT-proBNP results obtained in primary health care were significantly higher than by the hospital comparison method (bias ranged from 14.3-23.7%), whereas there was no significant bias when Cobas h 232 was used in the hospital setting (bias ranged from - 4.9 to 7.0%). User-friendliness of Cobas h 232 was overall acceptable. Conclusion. Cobas h 232 point-of-care instrument for measurement of NT-proBNP performed satisfactorily with regard to precision, user-friendliness, and lot-variation. A decrease in NT-proBNP levels observed in samples transported to a central laboratory needs further attention and investigation.

OriginalsprogEngelsk
BogserieScandinavian Journal of Clinical and Laboratory Investigation
Vol/bind75
Udgave nummer7
Sider (fra-til)602-609
Antal sider8
ISSN0036-5513
DOI
StatusUdgivet - 3 okt. 2015

Bibliografisk note

Publisher Copyright:
© 2015 Informa Healthcare.

Fingeraftryk

Dyk ned i forskningsemnerne om 'NT-proBNP on Cobas h 232 in point-of-care testing: Performance in the primary health care versus in the hospital laboratory'. Sammen danner de et unikt fingeraftryk.

Citationsformater