Biopsy-based optimization and calibration of a signal-intensity-ratio-based MRI method (1.5 Tesla) in a dextran-iron loaded mini-pig model, enabling estimation of very high liver iron concentrations

Peter D. Jensen*, Asbjørn H. Nielsen, Carsten W. Simonsen, Kenneth K. Jensen, Martin Bøgsted, Anne B. H. Jensen, Benedict Kjaergaard

*Kontaktforfatter

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

2 Citationer (Scopus)
42 Downloads (Pure)

Abstract

Objective: Magnetic resonance imaging (MRI)-based techniques for non-invasive assessing liver iron concentration (LIC) in patients with iron overload have a limited upper measuring range around 35 mg/g dry weight, caused by signal loss from accelerated T1-, T2-, T2* shortening with increasing LIC. Expansion of this range is necessary to allow evaluation of patients with very high LIC. Aim: To assess measuring range of a gradient-echo R2* method and a T1-weighted spin-echo (SE), signal intensity ratio (SIR)-based method (TE = 25 ms, TR = 560 ms), and to extend the upper measuring range of the SIR method by optimizing echo time (TE) and repetition time (TR) in iron-loaded minipigs. Methods: Thirteen mini pigs were followed up during dextran-iron loading with repeated percutaneous liver biopsies for chemical LIC measurement and MRIs for parallel non-invasive estimation of LIC (81 examinations) using different TEs and TRs. Results: SIR and R2* method had similar upper measuring range around 34 mg/g and similar method agreement. Using TE = 12 ms and TR = 1200 ms extended the upper measuring range to 115 mg/g and yielded good method of agreement. Discussion: The wider measuring range is likely caused by lesser sensitivity of the SE sequence to iron, due to shorter TE, leading to later signal loss at high LIC, allowing evaluation of most severe hepatic iron overload. Validation in iron-loaded patients is necessary.

OriginalsprogEngelsk
TidsskriftMagnetic Resonance Materials in Physics, Biology and Medicine
Vol/bind35
Udgave nummer5
Sider (fra-til)843-859
Antal sider17
ISSN0968-5243
DOI
StatusUdgivet - okt. 2022

Bibliografisk note

© 2022. The Author(s).

Fingeraftryk

Dyk ned i forskningsemnerne om 'Biopsy-based optimization and calibration of a signal-intensity-ratio-based MRI method (1.5 Tesla) in a dextran-iron loaded mini-pig model, enabling estimation of very high liver iron concentrations'. Sammen danner de et unikt fingeraftryk.

Citationsformater